Description
The DH-2000 light source offers stable, continuous output from 215-2500 nm. Using a combination of deuterium and halogen lamps, the DH-2000 is ideal for measuring a sample that has multiple features in different spectral regions or for analyzing a variety of different samples in your lab. All models of the DH-2000 family include easy-to-replace bulbs, keeping your measurements going with fast lamp changes.
For those that need light into the deep UV, the DH-2000-FHS-DUV-TTL and DH-2000-S-DUV-TTL models include a special deep-UV deuterium lamp providing output from 190-2500 nm. Both models come with a shutter to limit the exposure of samples to the UV light.
The DH-2000-FHS-DUV-TTL comes with a filter holder for filters up to 25 mm square or 25 mm round in diameter and 4 mm in thickness, plus a shutter controlled via TTL signal or switch up to 5 Hz.
Powerful and stable — deuterium (25 W) and halogen lamps (20 W) combined in a single light source
Integrated shutter — helps protect light-sensitive samples
Easy-to-replace, pre-aligned bulbs — minimal downtime needed for bulb replacement
TTL control — adds another layer of operation flexibility
DH-2000-BAL | DH-2000-FHS-DUV-TTL | DH-2000-S-DUV-TTL | FLAME-DA-CUV-UV-VIS | |
Wavelength Range | 210nm - 2.5μm | 190nm - 2.5μm | 190nm - 2.5μm | 200nm - 1.1μm |
Source | Deuterium & Tungsten Halogen | Deep-UV Deuterium & Tungsten Halogen | Deep-UV Deuterium & Tungsten Halogen | Deuterium & Tungsten Halogen |
Nominal Bulb Power | 25 W (deuterium) 20 W (tungsten halogen) |
20 W (tungsten halogen) 26 W (deuterium) |
20 W (tungsten halogen) 26 W (deuterium) |
— |
Typical Output Power | 194 µW (deuterium bulb) 615 µW (tungsten bulb) |
217 μW (deuterium) 295 μW (tungsten halogen) |
585 μW (deuterium) 990 μW (tungsten halogen) |
— |
Source Lifetime | 1000 hours | 1000 hours | 1000 hours | — |
Stability of Light Source Output | ≤0.1%/hour @ 254 nm (deuterium) ≤0.1%/hour @ 700 nm (tungsten halogen) |
— | — | — |
Warm-Up Time | 25 minutes | 25 minutes | 25 minutes | — |
Safety & Regulatory | CE ROHS WEEE |
CE ROHS WEEE |
CE ROHS WEEE |
— |
Operating Humidity Specification | 5-95% without condensation at 40 °C | 5-95% without condensation at 40 °C | 5-95% without condensation at 40 °C | — |
Operating Temperature | 5 °C – 35 °C | 5 °C – 35 °C | 5 °C – 35 °C | — |
Shutter | Yes | — | Yes | — |
Fiber Connector | SMA 905 | SMA 905 | SMA 905 | — |
Cuvette Pathlength | — | — | — | 10 mm |
Dimensions | 15 x 13.5 x 28.5 cm | 15 x 13.5 x 28.5 cm | 15 x 13.5 x 28.5 cm | 88.9 x 124.1 x 40.1 mm (with feet) |
Integrated Filter Holder | No | No | No | Yes |
Power Consumption | Approximately 78VA | Approximately 78VA | Approximately 78VA | — |
Trigger/Shutter Input Connection | SUB-D-15 pin | — | SUB-D-15 pin | — |
Trigger/Shutter Input Signal | TTL Up to 2.5 Hz maximum |
— | TTL Up to 2.5 Hz maximum |
— |
Weight | 5.5 kg | 5 kg | 5 kg | 350 g |
Integrated Cuvette Holder | No | No | No | Yes |
Power Requirements | 85-264 V 50/60 Hz | 85-264 V 50/60 Hz | 85-264 V 50/60 Hz | 5V DC external power supply (included) |
Payment & Security
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.